Accurate™ Fentanyl 1 ng/mL Rapid Urine Cassette Drug Test from Healgen®, CLIA Waived, OTC, 25/Box
Brand : Accurate®
- SKU:
- GBFEN-102g
- Per Each:
- 25
Accurate™ Fentanyl 1 ng/mL, CLIA Waived, OTC, Rapid Urine Cassette Drug Test from Healgen® Scientific, 25/Box
A rapid, screening test for the qualitative detection of Fentanyl in human urine at a cutoff concentration of 1 ng/mL.
Features
- Detection Level at 1 ng/mL
- CLIA Waived
- FDA Cleared for Over-the-Counter Use (OTC)
- Results in 5 Miutes
- Easy Visual Interpretation
- Simple Collection and Testing Procedure
- Low Cross-Reactivity with the Metabolite Norfentanyl
High Specificity for Fentanyl
Reduce False Positives: Assays that do not cross-react with norfentanyl ensure that positive results are specifically due to the presence of fentanyl, not its metabolites. This is crucial for accurately identifying current fentanyl use rather than past use.
Improve Monitoring: Specific assays help distinguish between recent fentanyl administration and the presence of residual metabolites from earlier use, aiding in more precise interpretations.
Product Documentation
Accurate™ Fentanyl 1 ng/mL Rapid Urine Cassette Drug Test from Healgen® Package Insert (IFU)
Directions For Use
If refrigerated, allow the test, urine specimen and/or controls to reach room temperature [59-86°F (15-30°C)] prior to testing.
1. Remove the Test Cassette from the sealed pouch and use within one hour.
2. Label the device with patient or control identifiers. Place the test cassette on a clean and level surface. Hold the dropper vertically and transfer 3 full drops of urine to the specimen well (S) of the test cassette, and then start the timer. Avoid trapping air bubbles in the specimen well (S). See the illustration below.
3. Set a timer for 5 minutes.
4. Read the results at 5 minutes. Do not interpret the result before 5 minutes and after 10 minutes. See the illustration in the test schematic.
5. If preliminary positive results are observed, send the urine sample to the laboratory for confirmation testing